our Online Store has been opened, please visit www.LoggerOnline.com
KDS230 Infusion Withdrawal Pump
Ideal for applications requiring multiple syringes, the KDS230 is an adaptation of the KDS210. The pump operates exactly like a KDS210, but has been modified to hold up to 10 syringes.
Features
Multiple syringe holder:
One to ten syringes, 10 µl - 10 ml
One to six syringes, 20 ml - 60 ml
One to four syringes, 100 ml - 140 ml
Withdrawal then infusion
Backlit LCD display
Knob locks/unlocks drive block for effortless, drag-free adjustment
Simple menu-driven setup:
select syringe size from displayed table
dispense volume
dispense flow rate
Continuous dispense volume display
Preset volume control and automatic shut-off
Settings can be reviewed or changed during operation
Stall detection
Choice of unit selection
Last settings stored in permanent memory
Built-in RS-232C interface for computer "daisy chaining" up to 100 pumps
TTL interface for foot switch, timer, relay control, outputs for run indicator, valve control
Independent rate and volume settings for infusion and withdrawal
Multiple mode selection:
Infusion
Withdrawal
Infusion then withdrawal
Withdrawal then infusion
Continuous cycle
Optional keypad programmable feature
Specifications
Pump Type: Infusion / Withdrawal
Max. No. of Syringes Ten
Syringe Size 10 µl to 10 ml (10 syringes)
20 ml to 60ml (6 syringes)
100 ml to 140 ml (4 syringes)
Dimensions 11 x 9 x 5.5 in.
28 x 23 x 14 cm
Weight 9 lb (4 kg)
Linear Force 40 lb (18 kg)
Advance Per Microstep 0.165 micron (1/16 step)
Max Step Rate (1/2 Step) 1600/sec
Min Step Rate 1 step / 100 sec.
Accuracy ± < 1%
Reproducibility ± 0.1%
Audible Alarm (Optional) **